Decon Pete explains how to ensure you are meeting UK guidance requirements with your ultrasonic cleaners and what records to keep for the CQC.

Firstly, let’s clarify the difference between testing and validation.

Validation is a way of confirming that decontamination equipment is working in the way the manufacturer intends. Validation also demonstrates that the equipment complies with the CE-marking process, under the terms of the Medical Devices regulations (MDR); and that the equipment cleans all instruments reliably and consistently.

According to UK guidance, practitioners must have their equipment validated and serviced annually, by an accredited engineer.

Nowadays most service contracts offered will include an element of validation.

In addition, when you purchase decontamination equipment, there are certain tests that you, as the end user, must complete to ensure that the physical conditions required for decontamination and sterilisation, are always achieved.

Unit de-gas and water change

Every time that the unit fills with fresh water and detergent it is important to run a de-gas cycle before full use.

The fresh water and detergent will contain oxygen bubbles. These will interfere with the cavitation of the unit and inhibit its ability to clean effectively.

The water and detergent solution needs changing at the end of every clinical session, or more often if visibility dirty.

Protein residue tests

On a weekly basis, ultrasonic baths must undertake protein residue tests to determine if any residual protein remains on the dental instruments.

The bath will contain RO or distilled water, along with a cleaning solution – enzymatic or otherwise, as recommended by the manufacturer.

At the end of the cleaning cycle, take a random instrument from the bath, rinse it, ideally with RO or distilled water, then swab for protein residue. If the instrument passes this test, all instruments can moved to the next stage of reprocessing.

Visual inspection

Visually inspect every single instrument that goes through reprocessing under an illuminated magnifier, before it moves on to the next stage.

If any visible debris is found, those instruments must go back through the cleaning process.

Ultrasonic activity test

On a quarterly basis, there are two tests that need carrying out.

The first is a foil ablation test (or standard foil test), also known as the ultrasonic activity test. This test determines that every possible area of the ultrasonic bath has cavitation happening within it.

Once the foil is in place, the bath should run on a standard cycle, or a time determined by the manufacturer. Once completed, the foil is removed and blotted dry, and then inspected for cavitation. This will appear as little perforations in the foil.

Cross-referenced the foil with that of the previous quarter. Make sure the perforations are in the same place, or as close as they were previously.

If they are not, or there are no perforations, this is an indication that cavitation is not taking place in that part of the bath. Store these foil pieces in a logbook, for reference in the following quarter.

Cleaning efficacy test

The cleaning efficacy test is the second quarterly test and is also known as the soil test. This is a manufacturer-led test, and not a UK guidance requirement.

The cleaning efficacy test requires the use of a PCD (process challenge device) and CEI (chemical efficacy indicator). The latter test employs the use of an indicator, which manufacturers normally paint red.

This goes inside the PCD and then into the ultrasonic bath. After a predetermined time (manufacturer dictated), the indicator is removed and there should be no visible red left on it. This is an indication that the cleaning efficacy is satisfactory.

Some manufactures state this should be a monthly test, so always check first.

Safety checks

You should regularly carry out safety checks to ensure that there is no damage to the bath; the inner chamber is in good order, the lid closes properly and there is also no visible damage externally.

These checks need documenting and kept for a minimum of two years. This will satisfy CQC or local inspectorate requirements.


If you need help with anything decontamination related, then please get in touch with me at info@deconpete.co.uk or visit the website www.deconpete.co.uk.

Follow Dentistry.co.uk on Instagram to keep up with all the latest dental news and trends.

Catch the previous Decon Pete columns:

  • The role of UK decontamination and interpretation of the guidance
  • Importance of in-practice training
  • Autoclave testing and validation
  • The importance of water quality
  • Manual or automatic cleaning?

The post Decon Pete – how to ensure ultrasonic cleaners meet UK guidance requirements appeared first on Dentistry.co.uk.





Source link

Related Articles